Reviews
Endoscopic video camera EndoGlance® V AF provides an image of ULTRA HD 4K 3840 × 2160 Pixels and equipped with a super-rapid auto-focused 0,5 The optical resolution is more than one second 1600 TV lines and high-end video processing help the surgeon to examine the fabric in detail, monitor the position of the tool and monitor the exact manipulations.
Super-rapid auto-focused for 0,5 seconds
The doctor selects the area of observation and initiates automatic focus with a button on the headhead 0,5 Seconds allow for quick restoration of the clear image after moving the endoscope, changing the distance to the tissue or moving to another section of the operating field.
Less manual action between inspection sites: a surgeon or an assistant holding the camera restores sharpness by one push, while keeping a convenient grip on the head. There is no need to rotate the focus ring manually. No object with constant focus distance 28 mm designed to operate with compatible endoscopic optics.
Deepening sharpness
The extended depth of sharpness increases the range of distances where anatomical structures remain sharp. The doctor can also observe the details at different depths and compare the position of the instrument with surrounding tissues. This helps to adjust the focus less often when there are small changes in distance to the observation object.
High-end image processing
8-Nuclear heterogeneous 64-The bit graphic processor provides high-performance video signal processing. The accelerated processing of the luminance changes of the observation scene and the ability to adjust color saturation helps to adapt the image to the intervention conditions and preferences of the operating surgeon.
Visual delay less 70 milliseconds helps compare instrument movements with monitor images for dissection, coagulation and stitching.
Optimal resolution more 1600 TV line
High quality image of ULTRA HD 4K combined with the optical resolution of more than 1600 The TV line allows for detailed examination of tissue surface, small vessels and limits of anatomic structures. At endoscopic diagnostics, high detail helps to detect low-level contusions, including on areas with signs of inflammation or suspicion of neoplasm, and to select areas for further evaluation.
During the operation, the doctor is more comfortable with the position of the needle, the working parts of the tool and the comparison of the dyes of the fabric 50/60P Ensures smooth display of motion.
Color Transfer BT.2020 and 16 Color adjustment steps
Support for Color Space BT.2020 The larger shade transmission helps the doctor to distinguish between the colours of the tissue, the vascular drawing and the boundaries of anatomical structures.
16 Color adjustment steps allow for better alignment of color transmission to the surgeon's preferences and intervention tasks.
Backlight sensor
The backlight sensor provides increased sensitivity and improves the image with low lighting. This helps to maintain visibility of the parts when examining less illuminated areas of the operating field and fields with uneven light distribution.
E-Glance algorithms for visibility improvement
E-HLC reduces the effect of bright lights on the image.
E-BLC increases the visibility of the parts in the darked areas of the operating field.
E-smoke improves image visibility when smoke is present, complementing physical removal.
E-REDE helps correct the expression of the red component of the image.
E-detail increases the display of fine parts and vascular structures.
E-moire reduces muareous artifacts when working with appropriate endoscopy optics.
More about video signal processing — page «Endoscopy image processing algorithms E-Glance».
Ten pre-established regimes
The camera offers ten presets for minimally invasive surgery and diagnostics, including laparoscopy, thoracoscopy, hysteroscopy, urology, rhinoscopy and arthroscopy. Presets speed up image setup. The clinician can adjust the settings to match the optics and their preferences.
Three programmable head-head buttons
The three programmable videohead buttons can assign any function available on the camera menu. The doctor selects the most requested command and adjusts the control to his preferences and intervention tasks:
Select and incorporate any E-Glance intellectual algorithm;
The image is increased and reduced;
Launch and stop video recording;
Photo production;
and other camera menu functions.
The surgeon or assistant can sharpen, adjust the image and save the important stage of the operation directly from the head-head, without going to the front panel of the chamber block. This control reduces the number of additional actions and allows for the retention of a convenient camera grab.
Easy videohead with autoclavic ability
The special design and material of the head restraint reduce its weight by facilitating retention and management of the endoscope during prolonged interventions 280 cm gives freedom of movement within the working space.
The videohead has a degree of protection IPX8. It is provided with liquid disinfection and autoclapping techniques, and is treated in accordance with the means, regimes and conditions established by the operating manual.
Video equipment 4K, Full HD and analog video
The camera is equipped with digital and analog exits to connect compatible medical monitors, recording devices and image transmission equipment:
4K — HDMI 2.0, DisplayPort (DP), 4 × 12G-SDI;
Full HD — 3G-SDI, DVI-D;
Analog Video — S-Video, CVS.
Also provided is a fiber-optic exit to transmit the video signal 12G. The choice of interfaces facilitates the inclusion of the camera in the existing operating video system and the selection of compatible image display and transmission equipment.
Photos, videos and audio comments
The camera allows you to keep photos and videos on the inner storage device with volume 2 TB or external USB: Records can be used to document interventions, re-review individual stages and train professionals.
The microphone entry allows for recording of the doctor ' s audio comments and sound from the operating room. The explanation for the intervention stages is supplemented by the operation and training materials. For telecasts, the camera is used as part of a compatible video and sound transmission system.
Select the endoscopic system to form Combinable laparoscope and other endoscopy optics with standard oculars, taking into account the direction of intervention and compatibility of the components.
Technical specifications
| Type | Endoscopic video camera of the ultra-high precision ULTRA HD colour image with automatic focus |
| Maximum resolution, pixels | 3840 Horizontal, 2160 Vertical |
| Sensor | 1/2.8 CMOS |
| Visual delay, ms | 70 |
| Image sensor, unit. | 1 |
| Color Standard | BT.709 |
| Focus | Electronic automatic, push button |
| Processor Type | 8-Nuclear heterogenic processor 64 bat |
| Party ratio | 16:9 |
| Signal/Noise ratio, dB | 50 |
| Scanner Method | Progressive Distillation 50/60P |
| Minimum illumination | < 1 Luxe, equivalent F1.6 |
| Videohead buttons | Electronic automatic focus, freely programmable 3 units. |
| Optimal resolution, tv lines | more than 1600 |
| Indirect resolution 4K 3840×2160 Pixels | Yes |
| Objective | Permanent focus distance |
| Focus distance, mm | 28 |
| Head size, with optical adapter, D x V x S, mm | 125 x 55 x 50 |
| Clamp mechanism | Standard Occulus adapter |
| Camera head temperature,℃ | No more than 37 |
| Scope | Laparoscopy, Toracoscopy |
| Video equipment 4K | HDMI2.0, DP, 4x12G-SDI |
| FULL HD Videos | 3G-SDI, DVI-D |
| Videos | S-Video, CVBS |
| FULL HD Video | 3G-SDI, DVI |
| Other interfaces | Pedal, RS232, Microphone, LAN, 4xUSB Rear Panel, 1xUSB Front panel |
| Control panel | Activation/off. Video, photo, white balance, brightness, menu |
| User scenes | 10 I'm gonna get you a setup scene |
| Digital Zoom | 5x increase, 50 Steps; 0.5 decrease, 5 Steps |
| Level luminance adjustment, unit. | 10 |
| Image modification | Horizontal turn, Vertical turn, Mirror display |
| Documentation | Image JPG/BMP 3840x2160, Video 3840х2160, 30fps or 1920х1080, 60Hz Mpeg-4 H.264 |
| Image Functions | E-Glance algorithms |
| Storage system | In-line hard drive 2 Tb, External storage device |
| Inter-tool communication | EndoGlance Net |
| Automatic white-base installation by manual object | Yes |
| The display luminance control buttons on the control unit | Yes |
| Video image button on control | Yes |
| Photo image button on the control unit | Yes |
| The digital video signal recorder on the flash drive is stored in the camera housing | Yes |
| Support for external video and photo storage | Yes |
| External storage interface version | USB 3.0 |
| The Sound Commenting Function | Yes |
| Microphone connection | Yes |
| Camera head cable | 280 cm Highly flexible |
| Camera head handling | Liquid, ethylene oxide, plasma hydrogen peroxide |
| Head protection | IPX8 |
Supplied components
| Video signal processing unit EndoGlance® ULTRA HD AF | 1 pc. |
| 32355, Head of camera with constant focus distance and auto-focused function | 1 pc. |
| 20046, Power cord | 1 pc. |
| 5015-72, HDMI-HDMI cable | 1 pc. |
| 5015-73, HDMI-DVI cable | 1 pc. |
| 5015-70, BNC-BNC cable (12G SDI) | 1 pc. |
| Operating Manual | 1 pc. |
| Flash storage device 128 GB | 1 pc. |
Alternative names
26.60.12.119-00000372, Video camera endoscopy EndoGlance® ULTRA HD AF for ТУ 26.60.12-018-47086606-2023
The purchase of products starts with a manager ' s consultation. Please describe your tasks or send a list of the items selected, and we will help you to select the components and check the compatibility of the system.
Press "Ask KP" in the product card, write on endo@endo.ru Or call on the phone +7 (843) 207-02-10. Please specify the names or articles of the product, the quantity, organization details and contact details.
Estimated delivery date — from 2 up to 45 It depends on the availability of products, the ordering and the loading of production.
The manager will agree on a specific time frame for your order. If the equipment is required by a specific date, report it when you are in circulation.
Yes Endomedium Develops and produces medical equipment with 1991 Production by the GCO Endomedium+».
The site only contains our products. All the trademarks here are owned by the public Endomedium. Registration and other documents issued under the «Documents».
Yes, our equipment and tools that are required to be registered as medical items have registered documents.
Documents are available in The appropriate section of the site. If you need help with the search, give me the name, model or artiful of the product — a specialist will pick up the documents related to it.
Guarantee and post-insurance maintenance is performed by an authorized service centre Endomedium-Services, licensed for maintenance and repair of medical equipment.
Treatment and maintenance materials are available under the following headings «Service Centre».
Contact the service centre: service@endo.ru, +7 (843) 249-00-45. Please specify the model and serial number of the product, describe the malfunction and leave contact details for feedback. Please attach photographs or videos if possible.
The specialist will specify the course of action and the delivery of the product Requirements for the preparation of service equipment.