Processing of roller pomp and accessories
Material is a product Endomedium: 5111-08 — Arthroscopy pump АНС-01, 5111-09 — Hysteroppa АНЖГ-01, 5111-11 — Urompa. Control and accessories require a separate approach: the method allowed for a tube or a metal transferor may not be acceptable for another component.
How to use the table
Find the exact name of the node and check the chosen sterilization method. Please consider the notes for the break-up and the drying. The method resolution in the table does not set the full cycle: duration, preparation and other parameters are specified by the product.
In the table, the sterilization of the control unit, the supply cord, the pneumatic pedal and the Type I single filter by the specified methods is prohibited. Do not apply to them the modes for multiple sterilized accessories.
Permitted sterilisation methods
| Host |
Sterilization method |
|||
|---|---|---|---|---|
|
Autoclavation at 134°C |
Autoclavation at 121°C |
Gas sterilisation of EO |
Sterilization in a dry-cup closet |
|
| Control unit arthroscopy pump | PROHIBITION | PROHIBITION | PROHIBITION | PROHIBITION |
| Power cord | PROHIBITION | PROHIBITION | PROHIBITION | PROHIBITION |
|
Pressure-retardant tube (With two things) *Metal pieces shall be removed from the silicone tube |
SETTLEMENTED | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED |
| Pressure-retardant tube | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED |
| Trinket | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED |
|
A needle for the sterilized fluids |
SETTLEMENTED | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED |
| Air supply needle | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED |
| Pneumatic pedal | PROHIBITION | PROHIBITION | PROHIBITION | PROHIBITION |
|
Aspiration receptacle 2 Litrate *Including plastic converters (Transitors) containers |
PROHIBITION | SETTLEMENTED | SETTLEMENTED | PROHIBITION |
|
Abacterial, multiple, autoclavic, Type II *The inside surfaces of the filter shall be thoroughly dried after sterilization |
SETTLEMENTED | SETTLEMENTED | SETTLEMENTED | PROHIBITION |
| Disposable bacterial filter, type I | PROHIBITION | PROHIBITION | PROHIBITION | PROHIBITION |
|
Squeeze-Fix Scrap (Squis-Fix) *The clamps shall be removed from the silicone tube and remain in the open state |
SETTLEMENTED | SETTLEMENTED | SETTLEMENTED | PROHIBITION |
| Transitioner | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED |
| Sylicone tube set | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED | SETTLEMENTED |
Preparation of individual supplies
The metal connections are removed from the silicone tube before processing. The Squeeze-Fix clamps are removed and left open. The internal surfaces of the Type II multiple filter after sterilization are thoroughly dried.
If you need to specify how a particular product is treated, please contact Technical support Endomedium. Name, article and model of product.