Do you need some heat CO₂ to laparoscopy? Impact of the temperature of the insuffraction gas
Heated CO₂ to laparoscopy Sometimes considered as an additional technology that should reduce the cooling of the patient's body and reduce post-operative pain.
But there is an important question: Is the use warm CO₂ to clinically significant advantages over standard room temperature gas?
A clinical analysis shows that convincing evidence of the significant advantage of heated CO₂ Moreover, there is an important technical factor: Gas directly heated inside insufflator, may lose heat significantly on the standard supply line. This was directly demonstrated in the temperature pilot study CO₂ at different points of the supply system.
Why are they even suggesting heating CO₂?
When pneumoperitoneum is created CO₂ It enters the abdomen in a large volume. Because the gas temperature is usually lower than the body temperature, it can in theory contribute to the cooling of the patient.
On this basis, the idea of preheating was proposed CO₂.
Logic seems obvious: if gas is heated to temperature close to body temperature, it should cool down the patient.
However, the effectiveness of this technology requires that more than the temperature of the gas be considered at exit insufflator, but also gas temperature Just before he enters the patient's system.
And here comes the fundamental technical problem.
Gas inside heating insufflator Doesn't mean the heated gas is coming to the patient
Between insufflator and the patient is the supply line CO₂.
If the gas is heated to a temperature above ambient temperature, it shall, when passing through the tube, heat the ambient air through the walls of the highway.
So the temperature CO₂ is reduced as it moves from insufflator to the patient.
In other words:
Gas outlet temperature insufflator ≠ temperature of gas at patient inlet.
This is particularly important when using standard lines without a heating element of their own.
Pilot study: hot gas cools down in the motorway
This effect was directly investigated by Jacobs, Kiechle and Morrison in the work Carbon Dioxide Gas Heating Inside Laparoscopic Insufflators Has No Effect, published in JSLS in 2005 Year.
Authors compared five standard laparoscopic insufflator, One of which had a built-in heater CO₂. The temperature of the gas was measured at different points of the system directly at the exit insufflator and at the end of the line 10 feet (3 meter).
Wow insufflator Snowden Pencer built-in heating allowed temperature to rise CO₂ at the outlet of the apparatus maximum 35,4 °C.
However, after passing through the 3-The temperature of the gas returned to almost the temperature of the room −0,22 up to +1,10 °C.
The authors also found that even when the flow was 20 L/min CO₂ reached the patient almost at ambient temperature.
So the experiment shows directly: heating CO₂ inside the shell insufflator does not ensure that the temperature of the gas remains elevated by the time it reaches the patient, if a normal line without its own heating is used between the apparatus and the patient.
PubMed — Carbon Dioxide Gas Heating Inside Laparoscopic Insufflators Has No Effect
Why Gas Chilling?
The reason is the usual heat exchange.
Heated temperature CO₂ So the heat from gas to the environment passes through the walls of the highway.
The cooling effect:
- Gas outlet temperature insufflator;
- Operating temperature;
- length of line;
- diameter of tube;
- material and wall thickness;
- Flow rate CO₂;
- Time of gas passing through the line.
The Jacobs and co-authors experiment confirms that even with the built-in heater insufflator The elevated gas temperature did not remain at the end of the standard 3-of the railway.
Therefore declared temperature CO₂ at the outlet of the apparatus cannot be automatically considered as the temperature of the gas entering the abdominal cavity.
Heating at exit insufflator And heat just in front of the patient - different technologies
From an engineering point of view, there are two fundamentally different approaches.
The first option is to heat the gas inside insufflator.
Gas is heated inside the machine and then passed through the normal line to the patient.
In this case, some heat is lost by passing through the tube 3 The temperature of the heated gas returned to approximately the temperature of the room.
The second option is to keep the temperature of the gas directly in the highway.
A special heated line with an integrated electrical heating component may be used.
In such a case, the heating occurs directly along the path of the gas, thereby compensating for heat losses.
It is therefore essential to distinguish between:
Normal:
Insufflator → heating of gas . normal tube
Heated line:
Insufflator → Hot line ♪ temperature maintenance ♪ patient
The heated line solves the problem of heat loss
If the temperature is to be kept elevated CO₂ until he gets to the patient, one heater inside insufflator Not enough.
Compensation for heat losses directly in the highway.
This is why there are special systems in which the heating appliance is located directly within the gas supply line.
Such systems allow the temperature of the gas to be maintained during its movement towards the patient, as opposed to the regular tube.
It's not just a function anymore insufflator, a separate gas temperature control system, A special line and related consumables.
Heated lines are specialized consumables
Special heated lines shall be produced for individual supply systems CO₂ and may be one-off.
For example, the FDA documentation on WESCAP Dissable Heating Tubing Set indicates the use of a single heating tube that contains a heating element.
Therefore, keeping the temperature of the gas immediately ahead of the patient requires more than just insufflator with integrated heating device, a compatible heating line system.
This increases the requirements for operating equipment and creates additional costs for specialized consumables.
For the Russian market, the question of the actual accessibility of such highways needs to be assessed separately by specific producers and models The need for specialized consumables and potentially higher operating costs, Not to claim without further analysis that such items are completely absent from the Russian market.
Heated inside insufflator has limited practical value
The question of principle thus arises.
The manufacturer may declare temperature CO₂ Exit insufflator, For example 35–37 °C.
But if gas passes several metres through the normal highway after that, a large part of the thermal energy is transferred to the environment.
Experimental data show that the gas in the system is heated at the exit insufflator up to 35,4 °C, after passing 3-The distance between the distance and the distance between the two is the same as the distance between the two.
Therefore heating function itself inside insufflator doesn't mean the patient gets warmed up yet CO₂.
To assess the real effect, the gas temperature must be measured immediately before it reaches the patient.
Do you even need a hot one CO₂?
Even if the technical problem of heat loss from a heated highway is solved, the main question remains:
Does it heat up CO₂ clinically significant benefit to the patient?
Randomized study of Nguien and co-authors included 20 Patients who were executed laparoscopic In all cases, external heating was used, and in one group, heating was used CO₂.
Gas heating did reduce the temperature decrease inside the abdominal cavity 1,5 The internal temperature has increased by about one hour since the operation began 0,2 °C in the heated gas group and dropped by approximately 0,5 °C in the control group.
However No significant difference in body temperature, post-op pain or morphine consumption was found.
Heated CO₂ and post-operative pain
Possible reduction of post-operative pain is one of the arguments of heating supporters CO₂.
However, the results of the studies are mixed.
No statistically significant reduction in post-op pain or morphine requirement was found in the study of Nguien and the co-authors using heated gas.
Other clinical studies also prevent the reduction of post-operative pain from being a stable and reproducible heating advantage CO₂.
Therefore, the allegation «warm CO₂ Reduces post-operative pain» Cannot be considered as a established universal clinical fact.
PubMed — Perrin et al., Effect of CO₂ gas warming on pain after laparoscopic surgery
More importantly, the temperature of the gas or the temperature of the patient?
If the aim is to prevent hypothermia, a much more direct approach is to actively warm the patient himself.
During the operation, the body's thermal balance is influenced by:
- Anesthesia;
- Operating temperature;
- duration of intervention;
- area of open tissue;
- Infotherapy therapy;
- heat transfer through skin;
- Ventilation;
- Use of active heating methods.
In the study of Nguien and the co-authors, all patients used external heating and added heated CO₂ did not lead to a significant difference in basic body temperature.
This makes it possible to see active heating as a more direct way of controlling the patient's thermal state, while heating CO₂ is predominantly local and indirect.
What the meta-analysis of temperature studies shows?
The Metaanalysis of Dean and the co-sponsors assessed the impact of the warm CO₂ Intra-operative temperature.
The authors found that the use of warm gas could lead to statistically significant increases in body temperature compared to standard gas, and noted that the number of studies was limited and their quality and results varied.
Therefore, there is a certain physiological effect of gas temperature.
However, the existence of a small statistical difference in body temperature does not automatically imply a proven clinical need to equip each person insufflator heating system.
The temperature of the gas in the patient is not the indicator on the display
In practical terms, a simple question should be asked when evaluating the heating system:
What a temperature CO₂ Right before entering the abdomen?
If the temperature is measured only inside or immediately downstream of the apparatus, this does not reflect the actual gas temperature after passing the line.
The study by Jacobs and the co-authors illustrates this problem: gas temperature at the outlet of the heated insufflator I've reached 35,4 °C, But at the end 3-The gas was already at almost ambient temperature.
Therefore, the whole heating system must be taken into account for an objective evaluation of the efficiency of the heating system:
insufflator → line ~ connections ~ troacar ~ patient.
Conclusion
Heated CO₂ Directly inside insufflator Doesn't guarantee hot gas will reach the patient.
A pilot study of Jacobs, Kiechle and Morrison showed that CO₂, heated at exit insufflator up to 35,4 °C, after passing the standard 3-The authors concluded that it had been built into the system and that it had been cooled to almost the ambient temperature insufflator The heater does not provide a clinically significant increase in gas temperature in the abdomen.
PubMed — Jacobs, Kiechle, Morrison, 2005
To keep the temperature elevated CO₂ A special procedure is required before it enters the body Heated line with integrated heating element, Compensating heat losses on the way from insufflator to the patient.
But even in dealing with this technical problem, the question of clinical efficiency remains CO₂ may reduce cooling within the abdomen, but this does not necessarily reduce post-op pain or result in a significant increase in the patient's body temperature.
So it's heated CO₂ should not be seen as a binding function laparoscopic insufflator.
It is more important to consider parameters that directly affect the safety and quality of the pneumoperitoneum when selecting equipment:
- Intra-abortion pressure accuracy;
- flow stability CO₂;
- speed of the set pressure;
- Leak control;
- Security system;
- Reliability of equipment;
- Availability and cost of consumables.
Scientific sources
-
Jacobs V.R., Kiechle M., Morrison J.E.
Carbon Dioxide Gas Heating Inside Laparoscopic Insufflators Has No Effect.
JSLS. 2005;9(2):208–212. PMID: 15984712.
PubMed | Full text -
Nguyen N.T., Furdui G., Fleming N.W., et al.
Effect of heated and humidified carbon dioxide gas on core temperature and postoperative pain: a randomized trial.
Surgical Endoscopy. 2002;16(7):1050–1054.
PubMed -
Perrin G., Fletcher A., et al.
Effect of CO₂ gas warming on pain after laparoscopic surgery.
Surgical Endoscopy. 1999.
PubMed -
Dean M., Ramsay R., Heriot A., et al.
Warmed, humidified CO₂ insufflation benefits intraoperative core temperature during laparoscopic surgery: A meta-analysis.
ANZ Journal of Surgery. 2017;87(5):E128–E136.
PubMed -
Birch D.W., Dang J.T., Switzer N.J., et al.
Heated insufflation with or without humidification for laparoscopic abdominal surgery.
Cochrane Database of Systematic Reviews. 2016;10:CD007821.
PubMed | Cochrane